Calculating the Current Density and Electric Field in Human Head by Multichannel Transcranial Magnetic Stimulation
Artikel i vetenskaplig tidskrift, 2009

In this paper, we present a designed multichannel transcranial magnetic stimulation (mTMS) system with 40 small coils. The induced current density and electric field in realistic human head model by mTMS were calculated by impedance method and the results were compared with figure-of-eight coil. The dependence of stimulation focality on the coil current direction was investigated. Properly adjusting the input current phases can improve the magnitude of the induced fields in the brain.

brain

tissues

electric field

Current density

impedance method

coils

multichannel

transcranial magnetic stimulation (mTMS)

human motor cortex

Författare

M. Lu

Faculdade de Ciencias, Universidade de Lisboa

S. Ueno

Kyushu University

T. Thorlin

Sahlgrenska Universitetssjukhuset

Mikael Persson

Chalmers, Signaler och system, Signalbehandling och medicinsk teknik

IEEE Transactions on Magnetics

0018-9464 (ISSN)

Vol. 45 3 1662-1665

Ämneskategorier (SSIF 2011)

Elektroteknik och elektronik

DOI

10.1109/TMAG.2009.2012770

Mer information

Skapat

2017-10-08