Calculating the Current Density and Electric Field in Human Head by Multichannel Transcranial Magnetic Stimulation
Journal article, 2009

In this paper, we present a designed multichannel transcranial magnetic stimulation (mTMS) system with 40 small coils. The induced current density and electric field in realistic human head model by mTMS were calculated by impedance method and the results were compared with figure-of-eight coil. The dependence of stimulation focality on the coil current direction was investigated. Properly adjusting the input current phases can improve the magnitude of the induced fields in the brain.

brain

tissues

electric field

Current density

impedance method

coils

multichannel

transcranial magnetic stimulation (mTMS)

human motor cortex

Author

M. Lu

Faculdade de Ciencias, Universidade de Lisboa

S. Ueno

Kyushu University

T. Thorlin

Sahlgrenska Universitetssjukhuset

Mikael Persson

Chalmers, Signals and Systems, Signalbehandling och medicinsk teknik

IEEE Transactions on Magnetics

0018-9464 (ISSN)

Vol. 45 3 1662-1665

Subject Categories (SSIF 2011)

Electrical Engineering, Electronic Engineering, Information Engineering

DOI

10.1109/TMAG.2009.2012770

More information