Characterization of HPGe detectors using Computed Tomography
Journal article, 2015

Computed Tomography (CT) high resolution imaging have been used to investigate if there is a significant change in the crystal-to-window distance, i.e. the air gap thickness, in a small n-type detector cooled to 77 K, and in a medium sized p-type HPGe detector when cooled to 100 K. The findings were compared to detector dimension data made available by the manufacturer. The air gap thickness increased by (0.38 +/- 0.07) mm for the n-type detector and by (0.40 +/- 0.15) mm for the p-type detector when the detectors were cooled to 77 resp. 100 K compared to at room temperature. Monte Carlo calculations indicate that these differences have a significant impact on the efficiency in close geometries (< 5 cm). In the energy range of 40-700 keV with a source placed directly on endcap, the change in detector efficiency with temperature is 1.9-2.9% for the n-type detector and 0.3-2.1% for the p-type detector. The measured air gap thickness when cooling the detector was 1.1 mm thicker than manufacturer data for the n-type detector and 0.2 mm thicker for the p-type detector. In the energy range of 40-700 keV and with a source on endcap, this result in a change in detector efficiency of 5.2-7.1% for the n-type detector and 0.2-1.0% for the p-type detector, Le the detector efficiency is overestimated using data available by the manufacturer.

HPGe

Computed Tomography

Gespecor

Gamma spectrometry

Monte Carlo modelling

Semi-empirical calibration

Author

A. Hedman

Totalforsvarets forskningsinstitut

Umea universitet

J. B. Gogani

Totalforsvarets forskningsinstitut

M. Granstrom

Totalforsvarets forskningsinstitut

Karl-Åke L Johansson

Umea universitet

J. S. Andersson

Umea universitet

Henrik Ramebäck

Chalmers, Chemistry and Chemical Engineering, Energy and Material

Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment

0168-9002 (ISSN)

Vol. 785 21-25

Subject Categories (SSIF 2011)

Subatomic Physics

DOI

10.1016/j.nima.2015.02.041

More information

Latest update

8/24/2020